DETECTION DEEP ERROR
Cause: Presence error DEEP. Action: Check the connection of the Deep Discharge board. Verify the configuration on SCSS. If everything is OK, replace the faulty board.

Cause: Presence error DEEP. Action: Check the connection of the Deep Discharge board. Verify the configuration on SCSS. If everything is OK, replace the faulty board.
Cause: Communication fault between the measurement board and the display board. Action: Check the cables between the measurement board and the display board.
Cause: Backup battery for settings is faulty or disconnected. Action: Check the battery voltage (3.3V) or replace with a new battery on the measurement board EA0149. Check the date and time configuration. Notes: For ...
Read more >
Cause: SD card cluster error (ERROR Bad Cluster). Action: Check the board connection, or replace if necessary.
Cause: SD card error (ERROR No SDCard). Action: Check the board connection, or replace if necessary.
Cause: Sensor error I out of range (displayed on the main screen). No current in the circuit or offset measurement. Action: Check if the issue occurs in both charge and discharge modes. Check the ...