ADSE 740
The Pitot-Static tester for laboratories and workshops
The ADSE 740 is a complete high performance dual pressure Ps and Pt standalone test bench specially designed to be used in the workshop or in the laboratory to test and calibrate all air data equipment such as altimeters, vertical speed indicators, air speed indicators, MACH-meter, air data computers …) and sensors.
The high precision embedded sensors enable the ADSE 740 to be used as a pressure standard.
The user interface is programmed using Windows® and Labview®, with the data base managed using a spreadsheet for easy evaluation, management, statistics and presentation.
ADSE 740 FEATURES
• Complete self check enable
• High accuracy, high resolution
• RVSM compliant
• Programmable leak test
• Programmable flight envelope to protect
• equipment under test
• Most important flight values dispalyed
• simultaneously
• Programmable (password write protected)
• test schedules
• Select pressure units: hPa; psi ; inHg ; mmHg ; ft
• m ; kts ; km/h ; Mach number
• SCPI Command interface (RS232 by default)
• Computer connections: HDMI, USB port
• Ethernet
Designed for laboratory use
The ADSE 740 Pitot Static Tester is a high precision Pitot/Static Air Data Test Set for workshops and laboratories. It is designed to control and calibrate altimeters, rate of climb indicators, airspeed indicators, and air data computers. It is delivered in a 19″ 6U rack-mounted enclosure and includes the control computer, measurement systems and test security features. This guarantees you a complete Pitot Static test solution for all your needs.
Designed to simplify workshop maintenance activities
The ADSE 740 test set has been specially designed to simplify pressure tests on equipment to be fitted on the aircraft (altimeter, airspeed indicator, rate of climb indicator, air data computers, angle of attack …) in the workshop or as part of an entire avionic test rig. The user interface is programmed under Windows® and Labview®, with a database managed in a spreadsheet format for easy evaluation, management, statistics and presentation.
Flexible
The ADSE 740 provides excellent flexibility to conduct all your tests. It provides the ability to conduct independent or simultaneous Measurement or generation of air pressures (Ps =altitude channel) and total (Pt=air speed channel, MACH number) on independent channels. In the “Program” mode, the operator can choose to enter at each program step the actual data read on the unit under test (UUT), or to let the program run and to manually take the values to read on the equipment under test on the touch screen display on the front of the ADSE 740.
ADSE 740 SPECIFICATIONS
MEASUREMENT SPECIFICATIONS
Function | Range | Accuracy |
---|---|---|
Altitude | -2,300 to 60,000 ft | ± 3 ft at 0 ft |
Altitude rate | 100 to 6,000 ft/min | ± 10 ft or ± 1 % |
Airspeed (1) | 0 to 800 kts | ± 0.5 kt at 50 kts |
Airspeed Mach | 0 to 4.0 Mach | ± 0.0006 M at 0.8 M at 25,000 ft |
Static sensor (2) | 72 to 1,100 hPa | ± 0.13 hPa |
Pitot sensor (2) | 0 to 1,455 DhPa | ± 0.30 DhPa |
(1) Accuracy after Autozero function
(2) Different range & rate of sensors available on request
Measurement specifications in standard conditions
• On-going developments may cause a change in specifications without notice.
• Special developments may be considered on a case by case basis.
DELIVERY INCLUDED ACCESSORIES
• Pressure and vacuum pumps
• 3 m Ps, Pt & AoA tube with connector AN-4 female /RB03-M
• 2m AC 110/220 V cable
• RS 232 PC interface
• PC software for Windows OS
• Two nipples RB03-M / AN-4 M
GENERAL SPECIFICATIONS
Temperature Operating | Operating 10°C to 40°C (50°F to 104°F) Storage -20°C to +60°C (-4°F to 140°F) (Extanded Range on-demand) |
Power supply | 90/264 VAC from 50 to 400Hz |
Dimensions | 440 x 260 x 380mm (17.3 x 10.2 x 14.9 inch) 20kg (44 lbs) |
Screen | 10” Touch Screen LCD colour |
Calibration | Recommended period 12 months |
Ease of use | Windows human/machine interface Program script Easy programming of test reports |
Ease of maintenance | Modular design permitting ease of accessibility to mechanical assemblies and electronic components |
OPTIONS
• Range on Ps: 80,000 ft or 100,000 ft
• Range on Pt: 350 or 1,000 kts
• Altitude coder Option
• SCPI Option
• EPR (Engine Pressure Ratio) Option
• Cabin Pressure Option
• Outflow Valve Option